Medical Appliances / Cardiac Pacemakers





BAIKAL SC Implantable Telemetering Electric Pacemaker







The “BAIKAL SC” SSI-type Implantable Telemetering Electric Pacemaker is intended for treatment of atrioventricular heart blocks at Morgagni-Stokes syndrome and other cardiac rhythm disorders by electric stimulation of ventricles or auricles.

Adopts unipolar or bipolar electrodes with IS-1 standard tips (3.2 mm).

Programmable by the UNI programmers.

The Pacemaker offers the following functions:

  • protection against tachycardia
  • time programming with pulse repetition rate of 30 to 450 ppm
  • measurement of electrode impedance in a range from 100 to 3000 ohm

The Electric Pacemaker has reduced teardropshaped design.

Implantation of the Pacemaker is carried out in cardiologic centres or cardiosurgery departments.

Telemetering information:

  • model and serial number
  • programmable parameters of stimulation
  • battery condition
  • statistics: imposed or perceived events
  • impedance of electrode

The Pacemaker's operation mode can be altered prior or after implantation by reprogramming the following parameters:

Stimulation mode

AOO, AAI, AAT, VOO, VVI, VVT, OFF

Frequency of stimulating pulses, ppm

30 to 150 ppm (step 5 ppm)

Amplitude of stimulating pulses, V

0.60, 1.00, 1.35, 1.75, 2.10, 2.50, 2.90, 3.25, 3.65, 4.00, 4.40, 4.80, 5.15, 5.55, 5.90, 6.20

Duration of stimulating pulses, ms

122 to 976 ms (step 122 ms)

R-wave threshhold of sensitivity, mV

1.0, 2.0, 3.0, 4.0, 5.0, 6.0, 7.0, 8.0

P-wave threshhold of sensitivity, mV

0.5, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0

Refractory period, ms

200 to 500 ms (step 12 ms)

Hysteresis, ms

0 to 300 ms (step 50 ms)

Input - output operation mode

UNI (unipolar) or BI (bipolar)

Programming of VARIO

mode by amplitude and duration



specifications

print






Specifications






The Pacemaker is supplied with the following parameters of standard program: BAIKAL SC
Stimulation mode VVI
Input-output operation mode UNI
Frequency of stimulating pulses, ppm 70
Amplitude of stimulating pulses, V 3.65
Duration of stimulating pulses, ms 488
R-wave threshold of sensitivity, mV 3.0
Refractory period, ms 260
Hysteresis, ms 0
VARIO mode Off
Overall dimensions, mm, max 52 х 43 х7.5
Weight, g, max 23





Cardiac Pacemakers








"BIOPHON"-Line Devices










2006 © FGUP "Izhevsky mekhanichesky zavod"