The “BAIKAL SC” SSI-type Implantable Telemetering Electric Pacemaker is intended for treatment of atrioventricular
heart blocks at Morgagni-Stokes syndrome and other cardiac rhythm disorders by electric stimulation of ventricles or auricles.
Adopts unipolar or bipolar electrodes with IS-1 standard tips (3.2 mm).
Programmable by the UNI programmers.
The Pacemaker offers the following functions:
-
protection against tachycardia
-
time programming with pulse repetition rate of 30 to 450 ppm
-
measurement of electrode impedance in a range from 100 to 3000 ohm
The Electric Pacemaker has reduced teardropshaped design.
Implantation of the Pacemaker is carried out in cardiologic centres or cardiosurgery departments.
Telemetering information:
- model and serial number
- programmable parameters of stimulation
- battery condition
- statistics: imposed or perceived events
- impedance of electrode
The Pacemaker's operation mode can be altered prior or after implantation by reprogramming the following
parameters:
|
Stimulation mode |
AOO, AAI, AAT, VOO, VVI, VVT, OFF |
|
Frequency of stimulating pulses, ppm |
30 to 150 ppm (step 5 ppm) |
|
Amplitude of stimulating pulses, V |
0.60, 1.00, 1.35, 1.75, 2.10, 2.50, 2.90, 3.25, 3.65, 4.00, 4.40, 4.80, 5.15, 5.55, 5.90, 6.20 |
|
Duration of stimulating pulses, ms |
122 to 976 ms (step 122 ms) |
|
R-wave threshhold of sensitivity, mV |
1.0, 2.0, 3.0, 4.0, 5.0, 6.0, 7.0, 8.0 |
|
P-wave threshhold of sensitivity, mV |
0.5, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0 |
|
Refractory period, ms |
200 to 500 ms (step 12 ms) |
|
Hysteresis, ms |
0 to 300 ms (step 50 ms) |
|
Input - output operation mode |
UNI (unipolar) or BI (bipolar) |
|
Programming of VARIO |
mode by amplitude and duration |
|