The BAIKAL-332 SSI-type Single-chamber Implantable Programmable Telemetering Electric Pacemaker is intended
for treatment of atrioventricular heart blocks at Margagni-Stokes syndrome and other cardiac rhythms disorders by electric
stimulation of ventricles or auricles.
Pacemaker BAIKAL-332-IS-3,2 adopts unipolar ar bipolar electrodes with IS-1 standard tips (3.2 mm). Pacemaker
BAIKAL-332-5,2 adopts unipolar electrodes (5.2 mm).
Programmable by means of PROGREKS-05-IMZ programmer.
Telemetering Information:
- model and serial number
- programmable parameters of stimulation
- battery condition
Statistics:
- imposed events
- perceived events
Operation modes:
| Stimulation modes |
A00, AAI, AAT, V00, VVI, VVT, OFF |
| Input-output operation mode |
UNI/BI |
| Stimulation frequency, imp./min |
30…70…155 (5 imp./min steps) |
| Amplitude of stimulating pulses, ms |
0.4, 0.8, 1.1, 1.5, 1.9, 2.3, 2.6, 3.0, 3.4, 3.8, 4.1, 4.5, 4.9, 5.3, 5.6, 6.0, 6.8, 7.5,
8.3, 9.0 |
| Duration of stimulating pulses, ms |
122…488…1220 (122 ms steps) |
| Threshhold of sensitivity to R-wave, mV |
1.0, 2.0, 3.0, 4.0, 5.0, 6.0, 7.0, 8.0 |
| Threshhold of sensitivity to P-wave, mV |
0.6, 1.2, 1.8, 2.4, 3.0, 3.6, 4.2, 4.8 |
| Refractory period, ms |
242, 267, 293, 318, 343, 369, 394, 419, 445, 470 |
| Hysteresis, ms |
0…400 (25 ms steps) |
| Programming of VARIO mode |
by amplitude and duration (ON, OFF) |
|