Medical Appliances / Cardiac Pacemakers





BAIKAL-DR (in production since 2009, 2nd half year)







COMPACT BILOCULAR FREQUENCY-ADAPTIVE ELECTRIC PACEMAKER WITH AUTOMATIC STIMULATION MODE FUNCTION AND EXTENSIVE STATISTICS

The “BAIKAL-DR” DDDR-type Implantable Bilocular Telemetering Frequency-adaptive Electric Pacemaker is intended for treatment of cardiac rhythm disorders by electric stimulation of ventricles and atriums. The Electric Pacemaker includes an activity transducer (sensor control), allowing to increase frequency of stimulating impulses depending on physical activity of the patient.

Adopts unipolar or bipolar electrodes with IS-1 (3.2 mm) standard tips.

Programmable by means of UNI programmers.

The pacemaker implements the next functions:
- automatic switch of stimulation mode with constant analysis of activity within refractory periods;
- extensive statistics of the events’ array displayed as convenient histograms and protection against pacemaker-mediated tachycardias;
- time programming with pulse rate of 30 … 460 ppm;
- measurement of electrode impedance of 200 to 3000 ohm;
- extra stimulation mode with maximum amplitude and widht of stimulative pulse.

Telemetering information:

  • model and serial number ;
  • programmable parameters of stimulation;
  • battery condition;
  • electrode’s impedance;
  • event’s marker.

Statistics:

  • imposed events (A and V channels);
  • perceived events (A-V channels);
  • combination of imposed / perceived events;
  • switchovers of stimulation mode.

OPERATION MODES

Stimulation mode : DDD, DDDR, DDI, DDIR, DОО,DOOR, AОО,AOOR, AAI, AAIR, AAT, AATR, VОО, VOOR, VVI, VVIR, VVT, VVTR
Input - output operation mode Unipolar/Bipolar
Frequency of stimulating pulses, ppm 30…70…120 (5 ppm step)
Frequency hysteresis, ms ВЫКЛ., 50…300 (50 ms step)
Frequency hysteresis with detection, ms ON; OFF
Sensor frequency of stimulation, max, ppm 90…180 (5 ppm step)
Response time under sensor control, s 10…50 (10 s step)
Recovery time of stimulation frequency, s under sensor control 60…960 (60 s step)
Rate of frequency growth under sensor control 1…16 (step 1)
Amplitude of stimulating pulses (A-V channel), V 0,60; 1,00; 1,35; 1,75; 2,10; 2,50; 2,90; 3,25; 3,65; 4,00; 4,40; 4,80; 5,15; 5,55; 5,90; 6,30
Duration of stimulating pulses (A-V channel), mcs 122…366…976 (step 122 mcs)
A-channel threshold of sensitivity (P-wave), mV 0,5; 1,0; 1,5; 2,0; 2,5; 3,0; 3,5; 4,0
V-channel threshold of sensitivity (R-wave), mV 1,0; 2,0; 3,0; 4,0; 5,0; 6,0; 7,0; 8,0
Refractory period (A channel), ms 200…300…500 (20 ms step)
Refractory period (V channel), ms 200…260…500 (20 ms step)
A-V delay, ms 80…180…300 (10 ms step)
A-V delay hysteresis, ms ВЫКЛ.; 50; 50 with detection
Dynamic A-V delay OFF, small, average, large
P-V delay (change of AV delay upon detection of P-wave, ms 0or -30
Frequency of conduction, max, ppm 60…120…180 (5 ppm step)
Blind period of A channel, ms 50…150…300 (50 ms step)
Blind period of V channel, ms 30…50…60 (10 ms step)
Automatic mode switch ON; OFF
Mode switch Slow, fast
Frequency of switching the stimulation mode, ppm 90…180 (5 ppm step)
Detection of pacemaker-mediated tachycardias ON; OFF
Programming of VARIO mode by amplitude and duration ON; OFF

(bold-type marking is for STANDARD program data)

Development contractor: Joint-Stock Company “Cardics”, Moscow



specifications

print






Specifications






Overall dimensions, mm, max 52 х 47 х 9.5
Weight, g, max 24
Type of electrode connector IS-1 BI 3.2 mm
Service life in "STANDARD" mode min 6.8 years
Power source - fluorocarbon lithium battery 1,0 А*h





Cardiac Pacemakers